ELISA Analyzer EA-10LS features a single sampling channel, providing pipetting ranges from 10µl to 1000µl. Includes four standard optical filters for precise wavelength selection during analysis. This automated unit integrates two racks specifically designed to accommodate 1000µl tips for efficient workflow. Our analyzer offers both air pressure detection and clot detection functionalities for precise reagent handling.
Sampling ChannelIncorporates a position sensing system
High precision pipetting
Offers TIP detection
Built-in microplate reader
Includes alarm features
Sampling Channel | 1 channel |
Sample Racks | 6 × 20 push-pull sample rack |
Liquid Level Detection | Air pressure detection and clot detection |
TIP | 1000µl, 2 TIP racks |
Pipetting | 10µl to 1000µl |
Pipetting Precision | 100µl: CV ≤ 1%, Accuracy ≤ 2.5% 1000µl: CV ≤ 0.5 %, Accuracy ≤ 1 % |
Reagent Position | 2 Reagent racks |
Dilution Plate Position | 1 Dilution plate Position with incubation and shaking |
Washer | 1 unit |
Washing Head | 16-channel (32-pin), double row cleaning, hanging installation, with leakage |
Injection Uniformity | CV<1.5% |
Washing Residual | < 2µl |
Reading Channels | 8 independent photometric channels mono and dichromatic reading |
Absorbance Range | 0 to 4 Abs |
Linear Range | 0 to 3 Abs |
Repeatability | CV ≤ 1% |
Stability | ≤ ± 0.003Abs |
Spectral Range | 400 to700nm |
Optical Filters | 4 standard filters (405, 450, 492,630 nm) Total of 10 filters position |
Reading Accuracy | ≤ ± 0.02Abs from 0 to 1 OD ≤ ± 0.03Abs from 1 to 2 OD |
Sensitivity | ≥ 0.01mg/L |
Channel Difference | ≤ 0.02Abs |
Incubators | 2 shaking incubators |
Cover Position | 1 (2 covers) |
Heating Method | Dry heating with cover |
Shaking Function | Horizontal circular oscillation, 1 to 3 levels adjustable, amplitude 2.5mm |
Temperature Range | RT to 60℃ |
Temperature Uniformity | ± 0.5℃ |
Temperature Fluctuation | ± 0.5℃ |
Software | Above Windows 7 system |
LIS System | Bi-direction, support HL7 protocol |
Results | Absorbance and results viewed by software |
Disinfection | UV lamp |
Lighting | LED |
IAP Function | Firmware can be upgraded online at any time |
Temperature | 10 to 40 ℃ |
Humidity | 30 to 80 % |
Power | AC220V ± 10%,50/60Hz (Standard) 110V ± 10%, 50/60 Hz (Optional) |
Dimension (W× D ×H) | 870 × 700 × 885 mm |
Packing Dimension (W× D ×H) | 1018 × 828 × 1130 mm |
Net Weight | 160kg |
Gross Weight | 180kg |
Accessory No. | Accessory Name |
1 | Printer |
2 | Built-in scanner |
3 | Biosafety Protection |
Medical Diagnostics: Used for the identification of antibodies, antigens, and other biomarkers associated with infectious diseases, autoimmune disorders, and cancer.
Food and Beverage Industry: Utilized for the detection of foodborne pathogens, allergens, and contaminants in food products, ensuring food safety.
Drug Discovery and Development: Employed in drug screening by assessing the binding affinity, potency, and efficacy of potential therapeutic compounds.